Run Function (Patient Monitoring) in Normal Mode
Operator's Manual
6-15
To automatically set lower alarm limits, press the HOME
key to display the Main screen, then
perform the following navigation bar options:
In the User configuration menu, select
NEXT
to scroll to
LOWER ALARM LIMITS
and select
AUTO
. Select
NEXT
to scroll to the channel or channels to be changed, and
INCREASE
and
DECREASE
options to adjust the variance value or values. The options for percentage below
baseline are 5% to 30%.
When finished, press the HOME
key to exit the User configuration menu and return to the
Main screen. On the Main screen, the new alarm limit is displayed (numerical and graphical).
Note:
The INVOS™ system will remain in Auto lower alarm mode with the selection of a new patient. However,
the lower alarm limits when selecting a new patient are set to default=40 prior to baseline capture.
Note:
If the variance value or values are changed for a channel subsequent to setting the baseline, the lower
alarm limit for the corresponding channel will be automatically changed to reflect the change in the
variance value.
Manual Lower Alarm Limits
To manually change the lower alarm limits, press the HOME
key to display the Main screen,
then perform the following navigation bar options:
In the User configuration menu, select
NEXT
to scroll to
LOWER ALARM LIMITS
and select
MANUAL
. Use the
INCREASE
and
DECREASE
options to adjust the alarm limit. The lower alarm
limit range is 15–90.
BASELINE MENU
EVENT MARK
ALARM AUDIO ON/OFF
NEXT MENU
OUTPUT SELECT
USER CONFIGURATION
TIME SCALE
NEXT MENU
NEXT
PREVIOUS
OFF
ON
NEXT PREVIOUS
MANUAL
AUTO
BASELINE MENU
EVENT MARK
ALARM AUDIO ON/OFF
NEXT MENU
OUTPUT SELECT
USER CONFIGURATION
TIME SCALE
NEXT MENU
NEXT
PREVIOUS
OFF
ON
NEXT PREVIOUS
MANUAL
AUTO
NEXT
PREVIOUS
INCREASE
DECREASE
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......