Run Function (Patient Monitoring) in Normal Mode
Operator's Manual
6-23
In the User configuration menu, select
NEXT
to scroll to
COLOR CODED LABELS
. Select
YES
or
NO
.
When finished, press the HOME
key to exit the User configuration menu and return to the
Main screen.
6.5.11
Two- or Four-Channel Monitoring
When using just one preamplifier, specify two-channel monitoring. When the system is set to
two-channel monitoring, only channels 1 and 2 appear on the screen, and only the top preampli-
fier port (1 and 2) is active. To use two preamplifiers, you must specify four-channel monitoring
to enable all channels to appear on the screen.
Note:
The INVOS™ system must be restarted for this change to take place.
Note:
Separate sets of user configuration parameters are stored for two- and four-channel monitoring. Previous
settings for two-channel monitoring will resume when this configuration is selected, and vice versa.
Note:
For quicker access to parameters on page 2 of the User configuration menu, select
PREVIOUS
to wrap
around from the top to bottom of page 1, then select
NEXT
to access page 2.
To change the software version from two- to four-channel monitoring, press the HOME
key
to display the Main screen, then perform the following navigation bar options:
In the User configuration menu, select
NEXT
to scroll to
2 OR 4 CHANNELS
. Select the
INCREASE
and
DECREASE
options to choose the number of channels monitored and displayed. Press the
ON/OFF
key twice to restart the system. When the monitor has returned to Run function, the
number of channels selected should be displayed.
BASELINE MENU
EVENT MARK
ALARM AUDIO ON/OFF
NEXT MENU
OUTPUT SELECT
USER CONFIGURATION
TIME SCALE
NEXT MENU
NEXT
PREVIOUS
OFF
ON
NEXT
PREVIOUS
INCREASE
DECREASE
NEXT
PREVIOUS
YES
NO
NEXT
PREVIOUS
INCREASE
DECREASE
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......