Run Function (Patient Monitoring) in Normal Mode
Operator's Manual
6-29
There may be a delay of up to 5 seconds between the detection of an alarm condition and its pre-
sentation.
Battery Capacity Alarm Notification
Battery operation symbol/gauge
—Appears in the lower right corner above the naviga-
tion bar when the INVOS™ system is operating on battery power. When using A/C power, no
battery symbol appears. The symbol fill indicates approximate battery capacity. The “BATTERY
LOW” message and alarm tone accompanies the depleted battery/symbol gauge if the battery
power is critical. Plug into A/C power or turn on the master (mains) switch on the monitor back.
The battery has an approximate 20 minute operation capacity. Recharge time from 50% discharge
to full charge is approximately 20 hours and from 100% discharge to full charge is approximately
24 hours.
Alarm Tones
Medium-Priority Alarms
If audio alarms are not muted, when a medium-priority alarm condition is present, the monitor
issues a medium-priority alarm tone. The alarm tone is a burst of three short beeps on the same
pitch, repeated at 9 second intervals until the alarm condition is resolved. Medium-priority alarm
conditions are listed in
Low-Priority Alarms
If audio alarms are not muted and no medium-priority alarm condition is present, when a low-
priority alarm condition is present, the monitor issues a low-priority alarm tone. The alarm tone is
a single beep, repeated at 16 second intervals while the corresponding visual alarm is presented.
Low-priority conditions are listed in
.
Silencing Audible Alarms
To silence audible alarms, press the HOME
key to display the Main screen, then select the
ALARM AUDIO ON/OFF
option.
Caution:
Selecting the ALARM AUDIO ON/OFF option will silence the audible alarms permanently until
reactivated or until the INVOS™ system is rebooted.
A bell
appears in the upper right corner of the screen. To activate alarms again when channels
are silenced, select the
ALARM AUDIO ON/OFF
option from the Main screen.
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......