Clinical Studies
B-2
Operator's Manual
Results
:
Figure B-1.
rSO
2
and fSO
2
, 20 Subjects
One subject had a bias value of 37.2%, which corresponded with a very low signal quality index
number (SQI=1). The remaining 41 subjects had SQIs that started at 10 and remained greater than
four throughout the study.
Subject 206’s absolute bias data was excluded from the analysis due to the low signal quality, but
the trend data was left in as trending is not dependent on SQI.
Transitions 96 and 97 data; last 20 subjects; 179 data points. y=0.9133x–0.1755;
R
2
=0.9575; bias=0.26; standard deviation=2.9.
Table B-1.
Bias, Error, and Correlation for rSO
2
and fSO
2
Individual subjects (N=20)
mean
Range
Pooled (N=180) transi-
tions
Difference
1
(Bias)%
1.
Less subject 206
2.49
–6.47 to +11.78
0.26
Error (standard deviation)%
2.13
0.85 to 4.22
2.90
Correlation (r2)
0.947
0.823 to 0.990
0.958
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......