Installation
3-4
Operator's Manual
3.6
VGA Output Port
This port provides standard video signals for an external monitor (see section
3.7
Potential Equalization Connector
The potential equalization connector (grounding post) provides a redundant external connection
to earth ground if necessary.
3.8
Alarm Speaker
The alarm speaker provides an audio output signal to alert the operator to problems and alarms
that may occur while monitoring (see sections
, and
3.9
Cooling Fan
The cooling fan provides air circulation through the monitor. Cooling air enters through the vents
in the back of the monitor and exhausts through the fan. The fan does not require maintenance.
WARNING:
Fire hazard: Do not block airflow to the back of the INVOS™ monitor.
3.10
AC Input and Fuse
The AC input provides a connection for AC power. The INVOS™ system will operate on 100-240 V
AC/50 or 60Hz. The AC input has a master (mains) power switch that turns AC power to the
monitor ON and OFF. If this switch is in the OFF position, the monitor can only be turned on from
the internal backup battery.
Caution:
When the INVOS™ system is not in use, the master (mains) power switch must be turned on with
AC power connected in order to keep the battery charged. The battery has a nominal capacity of
1.2 Ah (ampere-hours). The charging time is 24 hours at 20°C for a fully discharged battery.
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......