Run Function (Patient Monitoring) in Normal Mode
Operator's Manual
6-25
Restore defaults does not affect the following settings:
•
Date
•
Time
•
Language
•
Number of channels (two or four)
•
Control mode
•
Intervention mode
•
Database features
•
Digital output enable
•
Digital output format
•
Alarm audio ON/OFF
•
Alarm audio off reminder
•
Settings for the channel count that is not currently selected
Note:
Changing the number of channels and restarting the monitor brings up the last saved user configuration
settings for the new number of channels. To restore defaults for both two-channel and four-channel
operation, change the number of channels, restart the monitor, and repeat the Restore defaults procedure.
6.5.13
Alarms and Status Messages
Alarms are visual and audible. Alarm messages and tones alert the INVOS™ system operator of a
condition in the patient or the INVOS™ system requiring attention. Upon alarm, the operator
should check the patient’s condition first to determine the cause of alarm. Sensor attachment to
the patient, upper and lower alarm limits, battery capacity, and other messages that may appear
on the screen should be verified to validate the alarm condition.
Visual communication of parameter values and alarms is visible by an operator with 20/20 vision
(corrected as necessary) from a distance of 1 meter when the monitor is placed in an environment
with ambient light in the range of 100 to 1500 lux and the operator’s position is anywhere on or
within the base of a cone subtended by an angle of 30° normal to the center of the plane of
display of the monitor. Visual communication of parameter alarm priority is visible under the same
conditions from 4 meters.
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......