Maintenance
9-4
Operator's Manual
9.6
Maintenance Schedule
9.7
Annual Inspection Procedure
Note:
The INVOS™ system should be inspected at least once every year for safety and normal operation.
9.7.1
Mechanical Inspection Points
Inspect the following for loose screws, cracks, dirt, loose connectors, corroded or bent pins, fluid
contamination or stains, or damaged cords.
•
Power cord
•
Reusable sensor cable or cables
•
Preamplifier or preamplifiers with cable or cables
•
Battery status (see section
)
•
Key panel
•
Fan
Table 9-1.
Maintenance Schedule
How often?
What to do
Annually
Check the INVOS™ system (see section
).
After 5 years
Replace the backup battery. (The battery is not user replaceable. See section
.)
After 5–6 years
Replace the lithium battery on the CPU. (The lithium battery is not user replaceable. See section
Who to Contact for Authorized Repair
.)
After 10 years
Replace the disk on the chip module. This replacement is only available at Covidien authorized
repair facilities. (See section
Who to Contact for Authorized Repair
As needed
LCD lamps (after 50 000 hours of operation)
Fan (after 40 000 hours of operation)
CPU (after 46 250 hours of operation)
These replacements are only available at Covidien authorized repair facilities. (See section
Who to Contact for Authorized Repair
.)
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......