Heat Manager Pro User Manual V1.5 Page 30
The data menu will appear:
[EXPORT DATA TO USB]
Write the existing buffer to an USB stick as a .csv file, together
with the entered batch data.
[SHOW NUMERIC DATA]
Numerical overview of the measurement buffer, if you want to see
the real values instead of the graph.
[DELETE DATA AND TRENDING] When you want the buffer cleared, push for 3 seconds on
the [DELETE DATA AND TRENDING] button. This operation is
irrevocable and all stored measurements are cleared.
Note:
this can only be pushed with all channels in stop.
[ENTER BATCH DATA]
Enter batch related information that will be stored under the a
separate filename with entered number, but with a different
extension
[SAMPLE-RATE (s.)]
Here you can adjust the sampling interval of the measurements.
This value is normally set to 150s, so every 150s the
Heatmanager Pro measures all thermocouples and stores its
values into the 4032 position measurement buffer
(last in first out, LIFO)
[RECORDER START/STOP] Start recorder with programmer or always on
[CSV AUTO WRITE]
Check & configure the auto write to the .CSV file on usb-stick or
sd-card (only with 4301TAD)