'Convert digits to displacement (inches)
Displacement = 3.598E-9 * Digits^2 + 1.202E-3 * Digits - 3.1682
'Convert temperature °C to °F
Temp_F = Temp_C * 1.8 + 32
NextScan
EndProg
8.1.5 Period Averaging — Details
Related Topics:
•
Period Average Measurements — Specifications
(p. 107)
•
Period Average Measurements — Overview
(p. 81)
•
Period Average Measurements — Details
(p. 465)
The CR6 can measure the period of a signal on a
U
terminal. See
Period
Average Measurements — Specifications
(p. 107)
for range, accuracy, and
resolution.
The measurement is performed as follows: low-level signals are amplified prior to
a voltage comparator. The internal voltage comparator is referenced to the
programmed threshold. The threshold parameter allows referencing the internal
voltage comparator to voltages other than 0 V. For example, a threshold of 2500
mV allows a 0 to 5 Vdc digital signal to be sensed by the internal comparator
without the need for additional input conditioning circuitry. The threshold allows
direct connection of standard digital signals, but it is not recommended for
small-amplitude sensor signals.
For sensor amplitudes less than 20 mV peak-to-peak, a dc blocking capacitor is
recommended to center the signal at CR6 ground (threshold = 0). Figure
Input
Conditioning Circuit for Period Averaging
(p. 466)
shows an example circuit.
A threshold other than zero results in offset voltage drift, limited accuracy
(≈
±10
mV
),
and limited resolution (≈1.2 mV).
The minimum pulse-width requirements increase (maximum frequency decreases)
with increasing gain. Signals larger than the specified maximum for a range will
saturate the gain stages and prevent operation up to the maximum specified
frequency. As shown in the schematics, back-to-back diodes are recommended
to limit large amplitude signals to within the input signal ranges.
Caution
Noisy signals with slow transitions through the voltage threshold
have the potential for extra counts around the comparator switch point. A
voltage comparator with 20 mV of hysteresis follows the voltage gain
stages. The effective input-referred hysteresis equals 20 mV divided by
the selected voltage gain. The effective input referred hysteresis on the
±
25 mV range is 2 mV; consequently, 2 mV of noise on the input signal
could cause extraneous counts. For best results, select the largest input
range (smallest gain) that meets the minimum input signal requirements.
Содержание CR6 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 36: ......
Страница 38: ......
Страница 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Страница 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Страница 136: ......
Страница 251: ...Section 7 Installation 251 FIGURE 46 Running Average Frequency Response FIGURE 47 Running Average Signal Attenuation ...
Страница 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Страница 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Страница 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Страница 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Страница 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Страница 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Страница 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Страница 542: ...Section 8 Operation 542 8 11 5 PCCard Memory Card Management FIGURE 128 CR1000KD PCCard Memory Card Management ...
Страница 546: ......
Страница 549: ...Section 9 Maintenance Details 549 FIGURE 133 Separate Back Shell from Module FIGURE 134 Disconnect Battery Connector ...
Страница 552: ......
Страница 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Страница 612: ......
Страница 648: ......
Страница 650: ......
Страница 688: ......
Страница 689: ......