Section 7. Installation
292
parameter. Subsequent measurements are scaled with the same multiplier.
FieldCal()
Option 3
does not affect
offset
.
Some measurement applications do not require determination of offset.
Frequency analysis, for example, may only require relative data to characterize
change.
Example Case: A soil-water sensor is to be used to detect a pulse of water moving
through soil. A pulse of soil water can be detected with an offset, but sensitivity
to the pulse is important, so an accurate multiplier is essential. To adjust the
sensitivity of the sensor, two soil samples, with volumetric water contents of 10%
and 35%, will provide two known points.
Calibration Report for Water Content Sensor
CRBasic Variable
At Deployment
SimulatedWaterContentSignal
175
mV
KnownWC
10
%
SimulatedWaterContentSignal
700
mV
KnownWC
35
%
WCMultiplier
0.0476
%/mV
The following procedure sets the sensitivity of a simulated soil water-content
sensor.
1. Send CRBasic example
FieldCal() Multiplier
(p. 292)
to the CR6.
2. To simulate the soil-water sensor signal, place a jumper wire between
terminals
U1
and
U11
.
3. Simulate deployment-calibration conditions in two stages as follows:
a. Set variable
SimulatedWaterContentSignal
to
175
. Set variable
KnownWC
to
10.0
.
b. Start the calibration by setting variable
CalMode
=
1
.
c. When
CalMode
increments to
3
, set variable
SimulatedWaterContentSignal
to
700
. Set variable
KnownWC
to
35
.
d. Resume the calibration by setting variable
CalMode
=
4
4. When variable
CalMode
increments to
6
, the calibration is complete.
Calibrated multiplier is
0.0476
.
Содержание CR6 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 36: ......
Страница 38: ......
Страница 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Страница 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Страница 136: ......
Страница 251: ...Section 7 Installation 251 FIGURE 46 Running Average Frequency Response FIGURE 47 Running Average Signal Attenuation ...
Страница 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Страница 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Страница 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Страница 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Страница 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Страница 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Страница 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Страница 542: ...Section 8 Operation 542 8 11 5 PCCard Memory Card Management FIGURE 128 CR1000KD PCCard Memory Card Management ...
Страница 546: ......
Страница 549: ...Section 9 Maintenance Details 549 FIGURE 133 Separate Back Shell from Module FIGURE 134 Disconnect Battery Connector ...
Страница 552: ......
Страница 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Страница 612: ......
Страница 648: ......
Страница 650: ......
Страница 688: ......
Страница 689: ......