Section 8. Operation
464
VSPECT Vibrating Wire Measurement
'This example program measures one Geokon 4450 vibrating wire displacement sensor. Sensor
'outputs are frequency (hertz) as a function of displacement and resistance (ohms) as a
'function of temperature. The VibratingWire() instruction outputs displacement measurement
'in terms of hertz and temperature in terms of degrees Celcius. After the VibratingWire()
'instruction, displacement is converted to inches of displacement, and temperature is
'converted to degrees Fahrenheit.
'Linear gage factors are taken from a Geokon 4450 calibration sheet. Temperature coefficients
'are industry standard for the thermistor type.
'Linear gage factors
'A = 3.598E-9
'B = 1.202E-3
'C = -3.1682
'Steinhart-Hart coefficients
'A = 1.4051E-3
'B = 2.369E-4
'C = 1.019E-7
'Wiring Diagram
'Geokon 4450 CR6
'----------- ---
'GROUND------------------|||-
'VW----------------------U1
'VW----------------------U2
'TEMP--------------------U3
'TEMP--------------------U4
Public
VWResults(6)
Alias
VWResults(1) = Hertz
Alias
VWResults(2) = Amplitude
Alias
VWResults(3) = SigToNoise
Alias
VWResults(4) = NoiseFreq
Alias
VWResults(5) = DecayRatio
Alias
VWResults(6) = Temp_C
Public
VWResults
Public
Displacement,Temp_F
Dim
Digits
BeginProg
Scan
(10,Sec,0,0)
'Measure displacement (hertz) and temperature (ohms)
VibratingWire
(VWResults(),0,0,1,U1,1000,2500,1,0.01,60,1.4051E-3,2.369E-4,1.019E-7)
'Convert hertz to digits
Digits = (Hertz/1000)^2 * 1000
Содержание CR6 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 36: ......
Страница 38: ......
Страница 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Страница 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Страница 136: ......
Страница 251: ...Section 7 Installation 251 FIGURE 46 Running Average Frequency Response FIGURE 47 Running Average Signal Attenuation ...
Страница 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Страница 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Страница 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Страница 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Страница 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Страница 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Страница 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Страница 542: ...Section 8 Operation 542 8 11 5 PCCard Memory Card Management FIGURE 128 CR1000KD PCCard Memory Card Management ...
Страница 546: ......
Страница 549: ...Section 9 Maintenance Details 549 FIGURE 133 Separate Back Shell from Module FIGURE 134 Disconnect Battery Connector ...
Страница 552: ......
Страница 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Страница 612: ......
Страница 648: ......
Страница 650: ......
Страница 688: ......
Страница 689: ......