Define (Method, Map, Formula and Curve)
|
67
ELx800 Operator’s Manual
Defining FORMULA
The ELx800™ supports three types of formulas (Cutoff, Transformation, and
Validation), as well as the ability to program variables for use within formulas. Up
to three types of Validation formulas may be defined (Blank, Control, and Assay
Validation).
Formula definition is not available in the EL800 version.
To define formulas:
1.
Start at the Main Menu and select
DEFINE
.
2.
Select the assay then press
ENTER
to display the
DEFINE
options screen.
D
E
F
I
N
E
:
M E T H O D M A P F O R M U L A C U R V E
3.
Select
FORMULA
. The
SELECT FORMULA TYPE
screen will appear
(described on the next page).
Calculation Structure
During data reduction, formulas are processed in the order shown below. The
number of permitted formulas of each type is shown as well.
•
Blank Validation
0-1
•
Control Validation
0-4
•
Assay Validation
0-4
•
Transformations 0-1
•
Cutoff Formulas
0-1
•
Curve-Fit Analysis (if a curve-fit method is defined)
To capture and manipulate the raw data using 384-well
microplates with the ELx800NB, you must use Gen5™, KC4™,
or KCjunior™. Gen5, KC4, and KCjunior are PC-based software
programs you can use to set up your assay, communicate with
the ELx800 to run the assay, and then manipulate the raw data
that is automatically retrieved from the reader. Refer to
Appendix B
for additional information.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...