Error Codes
|
143
ELx800 Operator's Manual
General Errors (Continued)
0601
Filter #1 Gain out of range
0602
Filter #2 Gain out of range
0603
Filter #3 Gain out of range
0604
Filter #4 Gain out of range
0605
Filter #5 Gain out of range
These errors indicate that the gain for the filter in question is out of the range
necessary to ensure performance to specifications.
Probable Causes:
•
Defective interference filter.
•
Defective lamp providing reduced signal, causing the error.
•
Misaligned optics.
0700
Reader failed noise test
Significant variations in background electronic noise were detected when
blocking the light and increasing the gain to maximum.
Probable Causes:
•
External signals may be penetrating the measurement chamber. Ensure
that the bottom and top shrouds, as well as the filter wheel cover, are
correctly installed.
•
A faulty motherboard (PN 7330400, 7330410, or 7330415) or daughter
board (PN 7330404 or 7330414).
•
Photodiode or power supply may be too close to the instrument.
•
Failure indicates excessive variation in the dark current (background)
noise levels of the measurement circuit.
0800
Reader failed offset test
During self-test, the background electronic signal detected is outside of
acceptable limits at maximum gain when blocking the light.
Probable Causes:
•
If noise Max is > 20000.
There may be an ambient light leak. Ensure that the plate carrier
door is properly closed.
Faulty motherboard (PN 7330400, 7330410, or 7330415) or daughter
board (PN 7330404, or 7330414).
Photodiode or power supply may be too close to the instrument.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...