![Biotek EL800 Скачать руководство пользователя страница 154](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074154.webp)
128
|
Chapter 5: Maintenance and Decontamination
BioTek Instruments, Inc.
Maintenance
A daily maintenance schedule is the best way to ensure accurate performance and a long
life for your instrument. Frequent cleaning of the microplate carrier and all exposed
surfaces of the instrument will help to reduce the amount of particulates or dust that can
cause erroneous readings.
Routine Cleaning Procedure
Warning! Internal Voltage.
Always turn off and
disconnect the reader from the power supply for all cleaning
operations.
Do not immerse the instrument, spray it with liquid, or use a
“wet” cloth. Do not allow the cleaning solution to run into the
interior of the instrument. If this happens, contact the BioTek
Service Department.
Do not soak the keypad—this will cause damage. Moisten a
clean cloth with deionized or distilled water and wipe the
keypad. Dry immediately with a clean, dry cloth.
Do not apply lubricants to the microplate carrier or carrier
track. Lubrication on the carrier mechanism or components in
the carrier compartment will attract dust and other particles,
which may obstruct the carrier path and cause the instrument
to produce an error.
Purpose
Exposed surfaces may be cleaned (not decontaminated) with a cloth moistened (not
soaked) with water or water and a mild detergent.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...