114
|
Chapter 4: Instrument Qualification
BioTek Instruments, Inc.
Liquid Testing
Conducting Liquid Tests confirms the ELx800’s ability to perform to specification with
liquid samples. Liquid testing differs from testing with the Absorbance Test Plate in
that liquid in the wells has a meniscus, whereas the test plate’s neutral density glass
filters do not. The optics characteristics may differ in these two cases, thus alerting the
operator to different types of problems. The liquid tests will help to detect optical
defects or contamination that can contribute to errant readings.
•
Liquid Test 1
confirms repeatability and consistency of the reader when a
solution is used in the microplate. If these tests pass, then the lens placement
and system cleanliness are proven.
•
Liquid Test 2
can be used to test the linearity, repeatability, and alignment of
the reader if you do
not
have an Absorbance Test Plate. Prepare the series of
solutions of varying absorbances as described on page 119.
•
Liquid Test 3
is an optional test, and can only be performed with the “UV”
model. It is offered for those sites that must have proof of linearity at
wavelengths lower than those attainable with the Absorbance Test Plate. This
test is optional since the ELx800™ has good “front end” linearity throughout
the specified wavelength range (see page 123).
•
BioTek offers a dye solution (PN 7120779, 25 ml; or 7120782, 125 ml) that may
be used in the stock solution formulation for Liquid Tests 1 and 2, or, if you
prefer, you may use the dye solution described in
Table 3
(page 119). The
purpose of the formulation is to create a solution that absorbs light at
~ 2.000 OD full strength when dispensed at 200
μ
l in a flat-bottom microplate
well.
•
Alternatively, any solution that gives a stable color will suffice. (This includes
substrates incubated with an enzyme preparation and then stopped with an
acidic or basic solution.) Some enzyme/substrate combinations that may be
used as alternates to the described dye are shown in
Table 2.
Table 2
Typical Enzyme-Substrate Combinations and Stopping Solutions
Enzyme
Substrate
Stopping Solution
Alkaline
Phosphate
o-nitrophenyl phosphate
3N sodium hydroxide
beta-
Galactosidase
o-nitrophenyl -beta-D galactopyranoside
1M sodium carbonate
Peroxidase 2,2'-Azino
di-ethylbenzothiazoline-sulfonic
acid (ABTS)
citrate-phosphate
buffer, pH 2.8
Peroxidase
o-phenylenediamine
0.03N sulfuric acid
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...