Specifications
|
7
ELx800 Operator's Manual
Ultraviolet/UV Model
The following specifications apply to 96-well, flat- or round-bottom plates.
Wavelength Range:
340 to 750 nm
Filters:
10 nm half-bandwidth interference filters.
User-accessible filter wheel. Up to 5 filters may
be installed on the instrument at one time.
Filters supplied: 340 nm, 405 nm, 450 nm,
490 nm and 630 nm.
•
Optical specifications
for the 400 to 750 nm range (
50-second
read
in normal read
mode):
Absorbance Measurement
Range:
0.000 to 3.000 OD
Accuracy:
± 1.0% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
Linearity:
± 1.0% from 0.000 to 2.000 OD @ 405 nm
± 3.0% from 2.000 to 3.000 OD @ 450 nm
Repeatability (STD):
± 0.5% ± 0.005 OD from 0.000 to 2.000 OD @ 405 nm
•
Optical specifications
for the 340 to 400 nm range
(50-second read
in normal read
mode
):
Absorbance Measurement
Range:
0.000 to 2.000 OD
Accuracy:
± 2.0% ± 0.010 OD from 0.000 to 2.000 OD
@ 340 nm
Linearity:
± 2.5% from 0.000 to 2.000 OD @ 340 nm
Repeatability (STD):
± 1.5% ± 0.005 OD from 0.000 to 2.000 OD
@ 340 nm
•
Optical specifications
for the 400 to 750 nm range
(30-second read
in rapid read
mode
):
Absorbance Measurement
Range:
0.000 to 3.000 OD
Accuracy:
± 2.0% ± 0.020 OD from 0.000 to 2.000 OD
@ 405 nm
Linearity:
± 2.0% from 0.000 to 2.000 OD @ 405 nm
Repeatability (STD):
± 1.0% ± 0.010 OD from 0.000 to 2.000 OD
@ 405 nm
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...