60
|
Chapter 3: Operation
BioTek Instruments, Inc.
Number of Standard Replicates
This option allows you to enter the number of replicates per standard group in
the assay. Any predefined value appears on the display.
E N T E R N U M B E R O F
S T A N D A R D R E P L I C A T E S :
0 2
•
Use the
numeric
keys to enter the
NUMBER OF STANDARD
REPLICATES
. The range is 1 to 8 replicates. The software will verify
that the number of replicates, multiplied by the number of standards,
does not exceed the number of wells on the plate.
•
Press
CLEAR
to clear the value on the display.
•
Press
ENTER
to continue.
Average Standards
The
AVERAGE STANDARDS
option allows you to select whether or not to
average the replicates of each standard group. This average is used to calculate
the standard curve instead of using the individual replicates of each standard.
If the number of standard replicates is 1, this option is not available.
A V E R A G E S T A N D A R D S ? Y E S
Y E S N O
•
Select
YES
to average the replicates for each standard group, and then
use the group averages when calculating the standard curve.
•
Select
NO
to use the individual standard replicates when calculating the
standard curve.
•
Press
ENTER
to continue.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...