![Biotek EL800 Скачать руководство пользователя страница 164](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074164.webp)
138
|
Chapter 6: Troubleshooting and Error Codes
BioTek Instruments, Inc.
Overview
If an error code appears on the display (or in Gen5™, KC4™, or KCjunior™), consult
the error code list to see if the error is easily correctable. If you cannot resolve the
problem, run the instrument
System Test
and note the exact error code and serial
number of the reader. Then call BioTek’s Technical Assistance Center (refer to
Chapter 1
for contact information).
System Test Description
The System Test feature conducts a series of tests, at each of five set wavelengths,
which confirm adequate light levels, low electronic noise, adequate photodiode
sensitivity, and overall system cleanliness. The testing is designed to ensure that the
ELx800™ will give in-specification performance for each set wavelength over the
specified OD range.
The reader automatically runs an internal System Test each time it is powered on. The
reader will “chirp” repeatedly if the power-on System Test results do not meet the
internally coded Failure Mode Effects Analysis (FMEA) criteria established by BioTek.
A system test should then be initiated to try to retrieve an error code from the reader.
Glossary of Terms
•
Air Blank:
A full light reading through a filter with no plate in the light
path.
•
Dark Current:
A reading taken with the light blocked to measure
background light levels in the reading chamber. Also used as a measure of
background electronic noise within the measurement circuit.
•
Gain:
An automatic electronic adjustment to the measurement circuit. The
gain adjustment compensates for changing light levels or filter variations.
For example, if the lamp output decreases slightly, the gain will increase to
make up the difference.
•
Axis:
Refers to a motor for the filter wheel or plate carrier.
•
Offset:
A numerical limit, usually a range. For example, if the gain fails an
offset test, it may be too high or too low.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...