Chapter 4: Instrument Qualification
BioTek Instruments, Inc.
Plate Preparation:
5
Pipette 200 µl of the concentrated solution from tube 1 into each well of
the first column, A1 to H1, of the microplate.
6
Pipette 200 µl from each of the remaining tubes into the wells of the
corresponding column of the microplate (tube 2 into wells A2 to H2,
etc.).
Note:
The choice of dilutions and the absorbance of the original
solution can be varied. Use
Table 3
as a model for calculating the
expected absorbances of a series of dilutions, given a different
absorbance of the original solution.
After pipetting the diluted test solution into the microplate and
before
reading the plate, we strongly recommend
shaking
the plate for four
minutes. This will allow any air bubbles in the solution to settle and the
meniscus to stabilize. If a plate shaker is not available, wait 20
minutes after pipetting the diluted test solution before reading the
plate.
Read Plate
1
Read the microplate prepared above five times using Normal Read
Mode dual wavelength at 450 nm with 630 nm as the blank.
Note:
Do not discard the plate; you will use it for the Alignment Test.
2
Print the raw data or export it to an Excel spreadsheet using Gen5™,
KC4™, or KCjunior™.
If you are exporting the data to an Excel spreadsheet, perform the
calculations described below and on the following page, and keep the
spreadsheet for future tests.
Linearity Test:
3
Calculate the mean absorbance for each well, and average the means for
each concentration.
4
Perform a regression analysis on the data to evaluate linearity.
For example, using Microsoft
®
Excel:
In a spreadsheet, create two columns labeled X and Y. Enter the actual
absorbance values in column X. Enter the expected absorbance values in
column Y.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...