Define (Method, Map, Formula and Curve)
|
61
ELx800 Operator’s Manual
Standard Concentrations
The Standard Concentration field allows you to enter the predicted or expected
concentration value for each standard group. If
manual mapping
was
selected, the replicate locations must also be defined.
C O N C N O F
L O C A T I O N
S T D 1 : 0 R E P # 1 : A 0 1
•
Use the
numeric
and
alpha
keys and the
decimal point
key to enter
standard concentration values. The range is 0.00001 to 999999. The entry
cannot exceed six characters including the decimal point.
•
If automatic mapping is selected, each replicate’s location is available
for viewing only. Pressing
ENTER
advances to the concentration value
entry for the next standard.
•
If manual mapping is selected, the location must be defined. Pressing
ENTER
from the standard concentration entry moves the cursor to the
LOCATION
field. Pressing
ENTER
from the
LOCATION
field advances to
the concentration value entry for the next standard.
Valid Well Locations
When defining the replicate locations, only the
alpha
keys are active for
the first character and
numeric
for the second and third characters. Valid
characters and numeric entries are based on the selected plate geometry.
The following table lists acceptable entries for well locations based on plate
geometry:
Plate Type
Range
6-Well
A01-B03
12-Well A01-C04
24-Well A01-D06
48-Well A01-F08
96-Well A01-H12
96H A01-H12
96M A01-H12
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...