Controlling the Reader With KC4™
|
153
ELx800 Operator's Manual
Problems
If KC4 fails to communicate with the reader and displays a serial communications
error, check the cable plug-in location to ensure that it matches the setup choices
and is not a Null cable. If this is suspected, add another Null and try again.
If an ‘Incorrect Reader Model Connected’
dialog is displayed, click
OK
to clear the
screen and select
System|Readers|Available Readers.
Verify that the reader
selected is correct.
Getting Started With KC4™
The following instructions
briefly
describe how to read a plate using KC4. Refer to
KC4's Help system and User's Guide early and often to learn how to create
protocols, assign well identifiers, read plates, print reports, and more.
To read a plate using KC4:
1.
Select
Data|New Plate.
2.
If prompted to select a protocol, select
Empty Protocol
and click
OK
. If not
prompted, select
Protocol|New,
or use KC4’s Protocol Wizard to step
through protocol creation.
3.
Select
Protocol|Reading.
The ‘Reading’ parameters dialog will appear.
4.
Select a
Reading Type.
5.
Define the
Filters
(wavelengths) at which the plate will be read.
6.
Select a
Plate Type
from the drop-down list.
7.
Define other reading parameters as necessary. Click the
Help
button for
assistance.
8.
When complete, click
OK.
9.
Select
Data|Read Plate.
The ‘Plate Reading’ dialog will appear.
10.
Enter any comments, place the plate on the carrier, then click
START
READING
to begin the plate read.
The plate will be read and then the raw data results will display in KC4.
To analyze, manipulate, or print results,
Protocol
parameters should be
defined. Refer to KC4's Help system or User's Guide for instructions.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...