![Biotek EL800 Скачать руководство пользователя страница 101](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074101.webp)
Define (Method, Map, Formula and Curve)
|
75
ELx800 Operator’s Manual
On paper, the formula reads:
(OD-(NC;x-CTL1;x))/(NC;x-CTL1;x)*100
On the reader, the formula
(NC;x-CTL1;x)
will be programmed as the
TVAR, since the transformation will apply to all standards, controls,
and samples on the plate.
At the
SCOPE VARIABLE
selection screen, choose
OD
and press
ENTER
.
Enter the formula
(NC;x-CTL1;x)
by using the
MATH
,
OTHER,
MAP
and
FUNCTION
keys. Press
ENTER
.
The formula definition screen is displayed. Choose
TRANS
.
Enter the formula
(OD-(TVAR)/(TVAR)*100)
using the
MATH
,
OTHER
,
MAP,
and
FUNCTION
keys. (“TVAR” is available as MAP
option.)
Example:
In the case of competitive reactions, converting absorbance data to
percent B/B
0
can be:
(OD/STD1)*100
. This divides all the wells by
STD1
,
presumably the 0 standard, and multiplies the results
by 100.
To do this:
At the
SCOPE VARIABLE
selection screen, choose
OD
and press
ENTER
.
Enter
STD1
as the TVAR formula. Press
ENTER
.
The formula selection screen is displayed. Choose
TRANS
.
Enter the formula
(OD/TVAR)*100
using the
MATH
,
OTHER
,
MAP,
and
FUNCTION
keys. “TVAR” is available as a MAP option.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...