![Biotek EL800 Скачать руководство пользователя страница 89](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074089.webp)
Define (Method, Map, Formula and Curve)
|
63
ELx800 Operator’s Manual
S T A N D A R D S O N P L A T E ? N O
Y
E
S
N
O
If
YES
is chosen, a new standard curve will be generated. The plate map is not
changed. (If “Prompt for Sample ID” is enabled in the UTIL section, you will be
prompted to enter the number of samples. See
Using the Utility Options
in
this chapter for more information on the UTIL options.)
If
NO
is chosen, the stored standard curve will be used. If
Auto mapping
had
been used to originally map the standards, blanks, controls and samples
defined for this assay, the map will be automatically regenerated without the
standards, beginning in well xxx (where xxx was chosen as the starting well in
the map, usually well A01). If
manual mapping
was used to map the plate,
the map is
not
regenerated – the reader will
not
produce results for the well
positions that originally were standards. Auto mapping is recommended, if
the standard’s curves will be routinely reused.
Number of Controls
The
NUMBER OF CONTROLS
screen allows you to enter the number of control
groups
that will be used in the assay. Any previously defined value will
appear on the display.
E N T E R N U M B E R O F
C
O
N
T
R
O
L
S :
0 2
•
Use the
numeric
keys to enter the
NUMBER OF CONTROLS
groups
in
the assay. For example, if the assay requires one or more positive
control wells and one or more negative control wells, enter 02.
•
The valid entry range depends on the number of locations on the plate
that are undefined. The maximum number of control groups is 8.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...