Chapter 3: Operation
BioTek Instruments, Inc.
Overview
IMPORTANT!
Do not turn on the instrument until the
carrier shipping straps and shipping block have been
removed.
The ELx800™ features a 25-pad keypad and a 2-line x 24-character LCD display, allowing
you to access the reader’s program menus and print test results. The reader’s bidirectional
serial port allows computer control of the instrument, and provides the means for
downloading additional assay definition files to the instrument. This chapter describes the
operation of the
open
(configurable)
assays onboard the ELx800
.
Recommendations for Achieving Optimum
Performance
Microplates should be perfectly clean and free of dust or bottom scratches. Use new
microplates from sealed packages. Do not allow dust to settle on the surface of the
solution; use microplate covers when not reading the plate. Filter solutions to remove
particulates that could cause erroneous readings.
Although the ELx800 supports standard flat, U-bottom, and V-bottom microplates,
optimum performance is achieved with optically clear, flat-bottomed wells.
Non-uniformity in the optical density of the well bottoms can cause loss of accuracy,
especially with U- and V-bottom polyvinyl microplates. Check for this by reading an
empty microplate. Dual-wavelength readings can eliminate this problem, or bring the
variation in density readings to within acceptable limits for most measurements.
Inaccuracy in pipetting has a large effect on measurements, especially if smaller volumes
of liquid are used. For best results, use at least 100
μ
l per well in a 96-well plate and 25
μ
l
in a 384-well plate.
Dispensing solution into 384-well plates often traps air bubbles in the wells, which may
result in inaccurate readings. A dual-wavelength reading method usually eliminates these
inaccuracies; however, for best results, remove the air bubbles by degassing the plate in a
vacuum chamber before reading.
The inclination of the meniscus can reduce reading accuracy in some solutions, especially
with small volumes. Agitate the microplate before reading to help bring this problem
within acceptable limits. Use Tween
®
20, if possible (or some other wetting agent) to
normalize the meniscus. Some solutions develop menisci over a period of several minutes.
This effect varies with the brand of microplate and the solution composition. As the center
of the meniscus drops and shortens the light path, the density readings change. The
meniscus shape will stabilize over time.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...