![Biotek EL800 Скачать руководство пользователя страница 213](http://html.mh-extra.com/html/biotek/el800/el800_operators-manual_2756074213.webp)
|
187
ELx800 Operator's Manual
Additional Materials and Equipment – Not Supplied
•
Automatic Microplate Plate Washer:
This is recommended; however,
plate washing can be performed manually.
•
Plate Reader:
Capable of measuring optical densities at 450 nm
referenced on air.
•
Distilled or Deionized Water:
This should be of the highest quality
available.
•
Calibrated Micropipettes:
For dispensing 1000, 100, and 10 µl.
•
Multichannel Pipette:
Recommended for dispensing 100 µl volumes of
conjugate, substrate, and stop solution.
•
Glass/Plastic Tubes:
For sample dilution.
Quality Control and Results
Quality Control
For an assay to be valid, all the following criteria must be met:
•
Standards and the positive and negative controls must be included in
each run.
•
The values obtained for all the controls should be in the ranges
specified on the QC Certificate.
•
The curve shape should be similar to the standard curve, shown on the
QC Certificate.
•
If the above criteria
are not met, the assay is invalid and the test should
be repeated.
Calculate Mean Optical Densities (for assays run in duplicate
only)
For each standard, control, and sample, calculate the mean OD of the duplicate
readings. The user must verify that the percentage coefficient of variation
(%CV) for each duplicate OD is less than 15.0%.
Содержание EL800
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Страница 2: ... ...
Страница 4: ... ...
Страница 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Страница 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Страница 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Страница 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Страница 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Страница 196: ...170 Appendix C Report Format BioTek Instruments Inc Figure 38 Samples with calls on Matrix Report ...
Страница 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Страница 198: ...172 Appendix C Report Format BioTek Instruments Inc Figure 40 Samples with calls on Column Report ...
Страница 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Страница 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Страница 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Страница 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Страница 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...