Maintenance
6.1 General
The BACTEC® 9050 instrument requires minimal maintenance from the user to
provide reliable performance. Daily verifications include: checking the incubator tem-
perature and printer paper supply (if configured). Routine preventive maintenance
consists only of a monthly changing or cleaning of the cabinet air filters.All other
procedures are on an "as needed" basis.Any maintenance or repair not described in
this section should be performed by Becton Dickinson personnel only.
Note that if testing is interrupted for longer than forty minutes in performing
any maintenance procedure. to ensure maximum recovery it is recommended
that you perform a subculture on ongoing culture vials.
MA-0103-5
53