7 - Troubleshooting
E01 Te m p e r a t u r e alarm
POSSIBLE CAUSE(S)
incubator temperature is too high (00000001) or too low (00000002).
Door was kept open too long.
Room temperature is not within recommended range.
• A i r filters are dirty, restricting fresh air intake.
• R T D , heater, fan, or blower is defective.
CORRECTIVE ACTION(S)
O C h e c k current incubator temperature on LCD Display to see if tem-
perature is still too high or too low.
O M i n i m i z e number and duration of door openings.
O M a k e sure room temperature is within range specified in Section 2 —
Installation and Setup.
O C l e a n or replace air filters.
O C h e c k thermometer vial to see if manual reading agrees with displayed
reading.
O R e s e t manual thermostat (see Figure 43 for location of Reset button).
O 00000001 —At temperatures in excess of 38° C, viability of many organ-
isms may have been lost. Recollection of specimens should be considered.
O 0 0 0 0 0 0 0 2 —At temperatures below 32° C, detection of some organisms
may be missed or delayed. Subculture of vials should be considered.
E02 R o t o r RPM out of
spec
POSSIBLE CAUSE(S)
Rotor speed is too fast, too slow (00000002), or stopped (00000004).
•
•
Something is impeding or jamming movement of rotor.
A vial is not seated in the station.
CORRECTIVE ACTION(S)
O C h e c k for and remove anything that may be impeding or jamming the
movement of the rotor.
O M a k e sure all vials are seated in stations.
O A l i vials should be subcultured.
MA-0103-E3
97