6 - Maintenance
6.2.4 Decontamination
A situation requiring biological decontamination of one o r more stations can
occur at any time.The priority in this situation is to first limit the extent of the
contamination and then to decontaminate the station o r replace the rotor if
complete decontamination cannot be accomplished. If there any doubts that the
affected station(s) cannot be completely decontaminated, the rotor should be
replaced.
To Decontaminate
a Station or Rotor
The solution recommended to clean the affected surface should be at least a I 0
per cent household bleach solution.All surfaces must be thoroughly washed with
the freshly prepared bleach solution so the surfaces are "glistening wet" If you
are not sure of the extent of the contamination, thoroughly wash the entire
exterior of the Rotor with the freshly prepared bleach solution. Proceed in the
following manner to decontaminate a station or Rotor if a vial should break in a
station.
1 W e a r gloves and a gown, completely covering any body surfaces that could
possibly come in contact with the affected instrument surfaces.
2 C a r e f u l l y remove broken glass with forceps, one piece at a time.
3 C o m p l e t e l y absorb the contaminated spill (gauze pads are most effective).
4 A p p l y the bleach solution to the affected surfaces so the surfaces are"glis-
tening wet."
5 A b s o r b the applied solution with gauze pads or paper towels.
6 T h o r o u g h l y dry all wet surfaces.
7 D i s c a r d ALL cleanup materials with biohazardous waste.
MA-0105-5
57