SACTEC' 9050 System User's Manual
4.8 Identifying Anonymous Vials
An anonymous vial is one which is physically in a station but has not been as-
signed to that station through
the
Vial Entry activity. It is very important that
when the system notifies you of anonymous vials (anonymous count in Summary
window increments and Identify Anonymous icon appears when door is opened),
you identify them as soon as possible.The anonymous state supersedes all oth-
ers, which means that vials that are anonymous may in fact also be positive, out-
of-protocol, in error, etc. However, before any other status can be displayed, the
anonymous vial must be identified to the system.
The Identify Anonymous function enables you to locate anonymous vials in the
instrument and assign them to stations.
To Identify Anonymous
Vials
Open the instrument door.
Follow the steps in Figure 15.
Note that because the removal and reinsertion of vials can result in an increase
of false positive rates, anonymous vial loading into the instrument should be
minimized.