Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
6 - Maintenance
6.3.3.4
Blower
The blower is located to the left of center facing the rear of the instrument. It is
a large cylindrical black module.To remove the blower, follow the steps below
and refer to Figure 38.
Required Materials
Small Phillips Screwdriver
To Replace
the Blower
1 R e m o v e the rear cabinet shell as described in Section 6.3.3. 1 .
2 U n p l u g the large black cable from the bulkhead fitting.
3 R e m o v e the four small phillips screws that hold the blower assembly in
place and remove the blower.
4 T o replace the blower, reverse the above steps.
Figure
3 8
- slower
Removal
MA-0103-D
79
Artisan Technology Group - Quality Instrumentation ... Guaranteed I (888) 88-SOURCE I www.artisantg.com