7 - Troubleshooting
E 0 9 N o t e s t s in over 40 minutes
POSSIBLE CAUSE(S)
• I n s t r u m e n t has been off or door has been opened for longer than 40
minutes.
• F o u r consecutive test cycles were missed (e.g., due to door openings).
• S y s t e m clock was set more than 40 minutes ahead.
CORRECTIVE ACTION(S)
• I f instrument has been off, four consecutive readings were missed, or
door has been open for longer than 40 minutes,
all vials should be
subcultured.
If system clock was set more than 40 minutes ahead,
subculture not required. If this error occurs and none of the above
events happened, write data to disk and contact Becton Dickinson.
El0 D a t a b a s e corruption
POSSIBLE CAUSE(S)
• D a t a b a s e checksum test failed.
CORRECTIVE ACTION(S)
O W r i t e data to disk and call Becton Dickinson.
O 00000002, 00000008 -All vials should be subcultured.
Ell P r i n t e r error
POSSIBLE CAUSE(S)
• P r i n t e r paper is jammed or exhausted.
• P r i n t e r cable is disconnected during printing.
• P r i n t e r power is turned off during printing.
• P r i n t e r was taken offline during printing.
CORRECTIVE ACTION(S)
I C h e c k paper and clear jam or add paper if necessary.
O C h e c k printer cables (power and communications), reattach if neces-
sary.
O T u r n printer power on.
o P l a c e printer online.
) R e q u e s t report again. It will not resume printing automatically when
error condition is corrected.