Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
O
Installation and
Setup
2.1
General
Requirements
The BACTEC® 9050 instrument should be installed in an area that is free from
undue vibration, direct sunlight, high humidity, dust, temperature extremes, and
corrosive or explosive vapors or gases.The system will operate within specifica-
tions in room temperatures between 18.3° — 32.0° C (65° — 89.6°
F).
Relative
humidity should be between 10% and 90% (non-condensing). Clearances on all
sides should be at least 12 inches. Environments which
exceed
these limits could
adversely effect the performance of the system components.
The incubator should maintain its temperature to within plus or minus 1.5° C of
the temperature setting (35° C).This accuracy can be assured only if the room
temperature meets the requirements given above.
M A - 0 1 ) 7 , :