BACTEC°
9050
System User's Manual
E05 Te m p e r a t u r e standardization error
The temperature standards are out of tolerance.
POSSIBLE CAUSE(S)
• Te m p e r a t u r e standards are not seated.
• C o v e r over temperature standards is not in place.
• D e b r i s , spilled media, or blood inside instrument is affecting tempera-
ture standard readings
CORRECTIVE ACTION(S)
O Check/reseat temperature standards.
O Check/adjust cover over temperature standards.
D C h e c k that there is no debris, spilled media, or blood inside instru-
ment. Clean if necessary.
E 0 6 R o t o r configuration
error
POSSIBLE CAUSE(S)
• R o t o r is not mounted correctly.
• H o l d - d o w n nut (hubcap) is not tight.
CORRECTIVE ACTION(S)
O C h e c k rotor mounting - correct if necessary.
O C h e c k hubcap and tighten if necessary.
E07 P o w e r supplies high/low
POSSIBLE CAUSE(S)
• Te m p o r a r y electrical anomaly.
CORRECTIVE ACTION(S)
O R e b o o t instrument.
98
MA-0103-5