To adjust the test protocol duration for Mycosis/IC—F medium (not available for
use in the USA) press the "move to other field" soft key to highlight that field.
Use the UP ARROW or DOWN ARROW key to increase or decrease the dis-
played value.You can choose from 5
to
42 days.The default value is 14 days.
To adjust the test protocol duration for Myco/F Lytic medium, press the "move
to other field" soft key to highlight that field. Use the UP ARROW or DOWN
ARROW key to increase or decrease the displayed value.You can choose from 5
to 42 days.The initial default value is 42 days.The recommended testing protocol
for the following organisms are: 7 days for yeast, 30 days for fungi, and 42 days for
mycobacteria.
Press the "configuration" soft key to advance to the next setup display, or press
the "exit" soft key to exit configuration mode.
Time and Time Format (2/9)
23-01-94
12:2 4
01/23/99
12,24
When this display is first accessed, the minutes value is
highlighted.To adjust the minutes, use the UP ARROW or
DOWN ARROW key to increase or decrease the displayed
value.
To adjust the hours value, press the"move to other field" soft key to highlight the
hours field. Use the UP ARROW or DOWN ARROW key to increase or de-
crease the displayed value.
To adjust the time format, press the "format' soft key (shown above).The default
time format (colon separated) is shown. Continue to press the "format" soft key
to rotate among the format choices until the desired selection is shown.You can
choose among period ( . ) or comma ( , ) or colon ( ; ) separators.
Press the "configuration" soft key to advance to the next setup display, or press
the "exit" soft key to exit configuration mode.
Date and
Date Format (3/9)
98
II
23-.01-94
12:2 4
When this display is first accessed, the year value (at de-
.:::1123/94 f a u l t , the right field) is highlighted.To adjust the year, use
12,24 t h e UP ARROW or DOWN ARROW key to increase or
decrease the displayed value.
To adjust the day value (at default, the middle field), press the "move to other
field" soft key to highlight the day field. Use the UP ARROW or D O W N AR-
ROW key to increase or decrease the displayed value.
To adjust the month value (at default, the left field), press the "move to other
field" soft key t o highlight the month field. Use the UP ARROW or DOWN
ARROW key to increase or decrease the displayed value.
To adjust the date format, press the "format" soft key (shown above).The default
date format (MM/DD/YY) is shown. Continue to press the "format" soft key to
rotate among the format choices until the desired selection is shown.You can
choose from the following: