Table 25: Available measurement parameters
Option bar
Description
Good
Bad
DN +- peak
Displays signal peaks for the downstream transducer.
100-2300
<100 or >2300
Temperature
Displays the gas temperature (from 0/4-20 mA input).
N.A.
N.A.
Pressure
Displays the gas pressure (from 0/4-20 mA input).
N.A.
N.A.
Act Vol.
Displays actual volumetric flow.
N.A.
N.A.
Std vol.
Displays standard volumetric flow.
N.A.
N.A.
Up transit S
1
Displays skan transit time upstream.
N.A.
N.A.
Dn transit S
1
Displays skan transit time downstream.
N.A.
N.A.
Delta T S
1
Displays skan Delta T.
N.A.
N.A.
Up transit M
1
Displays measure transit time upstream.
N.A.
N.A.
Dn transit M
1
Displays measure transit time downstream.
N.A.
N.A.
Delta T M
1
Displays measure Delta T.
N.A.
N.A.
Vinst
Displays the instantaneous velocity.
N.A.
N.A.
1
available only if burst mode = S/M
NOTE:
The measurement units that appear in these
prompts are those selected in the global-system menu
earlier in this section.
5. Enter a flow rate value for the
zero
(low) end of the
analog output range and click [next item].
6. Enter a flow rate value for the
full
(high) end of the
analog output range and click [next item].
After responding to the above prompt, the meter returns
to step 1. Click [exit page] to proceed to another menu
or option.
Option card analog inputs
Complete the following steps to set up the analog inputs
of an option card installed in slot 1 (refer to
1. Double-click on the desired
input
(A, B, C or D).
NOTE:
The setup of input A is used as an example. Identical
procedures would be used to set up the remaining inputs.
2. Enter a
label
of up to eight characters for input A and
click on [next item].
3. Double-click on
off
to disable input A and return to step
1, or on
temp
or
spec
to designate it as a temperature or
special input.
•
Off
If OFF was selected to disable input A, the meter returns
to the
input
prompt in step 1. Double-click on another
input, or click [exit page] to move to another menu.
•
Temp or presr
If
temp
or
presr
was selected to set up input A as a
live temperature or pressure input, complete steps
4 and 5 below:
4. Enter a value for the
base
(low) end of the analog input
range and click [next item].
5. Enter a value for the
full scale
(high) end of the analog
input range and click [next item].
After responding to the above prompt, the meter returns
to step 1. Click [exit age] to proceed to another menu or
option.
•
Spec
If
spec
was selected to set up input A as a live special
input, complete steps 6 through 9 below:
6. Enter a
name
for input A and click [next item].
69
Содержание DigitalFlow XGF868i
Страница 1: ...DigitalFlow XGF868i Panametrics flare gas flow transmitter Programming manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page iv...
Страница 10: ...x...
Страница 38: ...28...
Страница 46: ...no content intended for this page 36...
Страница 54: ...44...
Страница 60: ...50...
Страница 88: ...78...
Страница 94: ...84...