Table 9: Available measurement parameters
Option bar Description
Good
Bad
PRESR
Displays the gas pressure (from 0/4-20 mA input).
N.A.
N.A.
Mw
Displays the molecular weight.
N.A.
N.A.
z
Displays the compressibility.
N.A.
N.A.
AcVOL
Displays actual volumetric flow.
N.A.
N.A.
StVOL
Displays standard volumetric flow.
N.A.
N.A.
Tu S1
Displays skan transit time upstream.
N.A.
N.A.
Td S
1
Displays skan transit time downstream.
N.A.
N.A.
DT S
1
Displays skan Delta T.
N.A.
N.A.
Tu M
1
Displays measure transit time upstream.
N.A.
N.A.
Td M
1
Displays measure transit time downstream.
N.A.
N.A.
DT M
1
Displays measure Delta T.
N.A.
N.A.
Vinst
Displays the instantaneous velocity.
N.A.
N.A.
1
available only if burst mode = S/M
NOTE:
The measurement units that appear in these prompts
are those selected in the
global-system
window earlier in
this section.
Procedure options
After completing the above steps, the XGF868i returns to the
I/O window. Do one of the following:
• To continue programming, refer to
, to navigate to the desired menu.
• To leave the
keypad program
, press [escape] three times.
1.9.2d Analog inputs
Refer to
, and complete the following
steps to set up the analog inputs of an option card installed
in slot 1:
Accessing the input
1. In the
keypad program
, scroll to
prog
and
press [enter].
2. In the
prog
menu, scroll to GLOBL and
press [enter].
3. In the global
program
menu, scroll to I/O and
press [enter].
4. Scroll to
optn
and press [enter].
5. Scroll to slot 1 and press [enter].
6. Scroll to the desired input and press [enter].
NOTE:
Input letters (A, B, etc.) correspond to input wiring
terminal numbers from top to bottom (e.g. input A
corresponds to pins 1,2 and 3; input B corresponds to
pins 4, 5 and 6; etc.).
Setting up the input
1. Enter a
label
of up to eight characters for input and
press [enter].
2. Scroll to the desired input
measurement
and
press [enter]. Do one of the following:
• If off was selected, go to
procedure options
below.
• If pressure or temperature was selected, proceed
to step 5.
• If special was selected to set up the input as a
live special input, proceed to the next step.
3. Use the arrow keys to enter a
name
for the input and
press [enter].
4. Use the arrow keys to enter a
unit
of measurement
for the input and press [enter].
16
Содержание DigitalFlow XGF868i
Страница 1: ...DigitalFlow XGF868i Panametrics flare gas flow transmitter Programming manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page iv...
Страница 10: ...x...
Страница 38: ...28...
Страница 46: ...no content intended for this page 36...
Страница 54: ...44...
Страница 60: ...50...
Страница 88: ...78...
Страница 94: ...84...