E.2 Configuration utility setup
The following is an
example
setup using National
Instruments Configuration Utility v3.1.
Figure 59
below shows the configuration utility with a
flowmeter on the network (Panametrics flow-XGF).
Figure 59: Configuration utility setup Example
NOTE:
The following procedures assume that the device
has been placed in the OOS (out-of-service) mode before
executing.
E.3 Selecting the desired
measurements
To set the measurement unit for each AI:
1. Double click on the FLOW transducer block (in the tree
under GEFlow-XGF).
2. Select the
others
tab and open the drop down list
for the PRIMARY_SELECTOR and SECONDARY_SELECTOR
(refer to
3. Choose the unit from the list (see
This unit will correspond to the unit that is available in the AI
block for network connection. The PRIMARY_SELECTOR unit
will correspond to ANALOG_INPUT_1 and the SECONDARY_
SELECTOR will correspond to ANALOG_INPUT_2.
4. After the desired measurements have been selected for
the PRIMARY and SECONDARY SELECTOR, choose the unit
system (UNIT_SELECTOR above the PRIMARY_SELECTOR)
that was programmed in the flowmeter (English or SI), as
shown in
Figure 60
below.
Figure 60: Primary selector drop down list
86