1.6 Entering transducer and pipe parameters
Enter the transducer and pipe parameters via the
pipe
submenu. While following the programming instructions,
refer to
1. From the channel
program
menu, scroll to the
pipe
option and press [enter].
2. The first prompt asks for the
transducer number
.
• For a standard transducer, use the arrow keys to
enter the number engraved on the transducer head,
and press [enter].
• If there is no number engraved on the transducer
head, press the right arrow key to scroll to the
STD
option, and use the up and down arrow keys to
change to
spec
. Then use the arrow keys to enter an
assigned number (from 91 to 99), and press [enter].
IMPORTANT
Special transducers have no engraved number on the
head and are rarely used. Examine the transducer head
carefully for a number.
• If you entered the number for a standard transducer,
proceed to the
pipe OD
prompt in step 5.
• If you entered the number for a special transducer,
proceed to step 3 below.
1.6.1 Special transducers
NOTE:
For special transducers, Panametrics supplies a
transducer data sheet with programming information.
3. Scroll to the transducer
frequency
(supplied by the
factory) and press [enter].
NOTE:
The frequency is required to transmit an excitation
voltage at the transducer’s natural frequency.
4. Enter the special transducer
tw
(time delay) value
(supplied by the factory) and press [enter].
Tw is the time required for the transducer signal to
travel through the transducer and its cable. This time
delay must be subtracted from the transit times of the
upstream and downstream transducers to ensure an
accurate flow rate measurement.
1.6.2 Pipe data
If either a standard or a special transducer is being used,
the programming sequence should be rejoined at this point.
5. To select the appropriate
pipe OD unit
type from
the list shown in
Table 3
, scroll to the right side of the
screen, and use the up and down arrow keys to step
through the list. Press [enter]. Then use the arrow
keys to enter the known pipe outside diameter or
circumference on the left side and press [enter].
Obtain the required information by measuring either
the pipe outside diameter (OD) or circumference at the
transducer installation site. The data may also be obtained
from standard pipe size tables found in
sound speeds
and
pipe size data
(914-004).
Table 3: Available pipe size units
English
Metric
Inches
mm = Millimeters
Feet
m = Meters
in/PI = Pipe circumference
in inches
mm/PI = Pipe circumference
in millimeters
ft/PI= Pipe circumference
in feet
m/PI = Pipe circumference
in meters
6. Use the arrow keys to enter the known
pipe wall thickness
(in inches or mm) and press [enter]. If the pipe wall
thickness is not available, look up the value in a table
of standard pipe size data which can be found in the
Sound speeds and pipe size data manual
(914-004)
IMPORTANT
Because the units cannot be independently chosen for
this parameter, the value must be entered in the same
units used for the pipe OD/circumference.
5
Содержание DigitalFlow XGF868i
Страница 1: ...DigitalFlow XGF868i Panametrics flare gas flow transmitter Programming manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page iv...
Страница 10: ...x...
Страница 38: ...28...
Страница 46: ...no content intended for this page 36...
Страница 54: ...44...
Страница 60: ...50...
Страница 88: ...78...
Страница 94: ...84...