
6-18
Chapter 6. Communication Data
Item
Data range
/ [unit]
Data address
Decimal
(hexadecimal)
R
W
Notes
(P-29)
Automatic display off
time
0 to 9999 [s]
2229 (08B5)
✓
✓
When set to “0,” the display is always
lit. (display turnoff function is
disabled)
(P-30)
PV display filter time
constant
0 to 9999 [ms]
2230 (08B6)
✓
✓
First-order lag low-pass filter for
display output
(P-31)
Keylock password
0 to 9999
2231 (08B7)
✓
✓
(P-32)
Analog scaling 2
(10 to 100 % FS)*
2
2232 (08B8)
✓
✓
The value is without a decimal
point.*
1
(P-33)
Totalization pulse
weight
[1/pulse]
2233 (08B9)
✓
✓
The value is without a decimal
point.*
1
(P-34)
Totalization pulse width
20 to 100 [ms]
2234 (08BA)
✓
✓
Set in multiples of 10.
*1. For the number of decimal places and the unit of data, see “Flow rate digits after the decimal point” (address 1003) and
“Flow rate unit” (address 1005).
Ex.: If the flow rate digits after the decimal point is set to “1” and the flow rate unit is set to “0” (mL/min), the value “1234”
means 123.4 mL/min.
*2. The setting range is the full scale flow rate (address 1002) multiplied by the percentage in parentheses. (The setting
range varies depending on the model and gas type.)
*3. If the data exceeds the range, the maximum value of the range is read.
*4. For the number of decimal places and the unit of data, see “Totalized flow rate digits after the decimal point” (address
1004) and “Totalized flow rate unit” (address 1006).
Содержание F4Q
Страница 10: ......
Страница 14: ......
Страница 46: ......
Страница 66: ...7 2 Chapter 7 Device Operation by Communication Commands...
Страница 68: ......
Страница 70: ......
Страница 71: ...Revision History CP SP 1458E Printed Edn New Page No Description June 2021 1...
Страница 72: ......
Страница 73: ......