____________________________________________________________________________
17
Output Regulation
±
(1
%
of setting
+
10 volts) from no load to full load
Maximum Capacitive Load 1.7 uF < 2 kV
0.8 uF < 4 kV
0.4 uF < 6 kV
0.3 uF < 8 kV
0.25 uF < 10 kV
0.2 uF < 12 kV
Discharge time
No load
≤
400 ms. Capacitive loading will increase the
discharge time. The time will vary due to the load
condition and voltage selection.
Dwell Timer
Range:
0.0, 0.4 – 999.9 seconds
0.0 - 999.9 minutes (0 = Constant)
Resolution:
0.1 second or minute increments
Accuracy:
±
(0.1% + 1 count)
Ramp-UP
Range:
0.4 – 999.9 sec
Ramp-DOWN
Range:
0.0, 1.0 – 999.9 sec (0=OFF)
Resolution:
0.1 sec increments
Accuracy:
±
(0.1% + 1 count)
Ground Fault
GFI Trip
Current:
1 mA max
Interrupt
HV Shut Down
response:
< 1 ms (Actual Output Voltage discharge
time is not included and vary due to load
condition)
3.3. 7715 Functional Specifications
INPUT
Voltage
115/230 VAC
±
10
%
, Single Phase, User selection
Frequency
50/60 Hz
±
5
%
Fuse
6.3 Amp, 250V Slow Blow
DIELECTRIC WITHSTAND TEST MODE
Output Rating
20 kV @ 10 mA
Output Adjustment
Range:
0.00 – 20.00 kV AC
Resolution:
10 volt/step
Accuracy:
±
(2
%
of setting
+
10 volts)
HI and LO-Limit
Range:
0.000 – 9.999 mA
Resolution:
0.001 mA/step
Accuracy:
±
(2
%
of setting
+
2 counts)
Arc Detection
Range:
1 – 9 at Output Voltage < 15.00 kV
1 – 7 at Output Voltage 15.00 kV
Failure Detector
Audible and Visual
Voltage Display
Range:
0.00 – 20.00 kV Full Scale
Resolution:
10 volt
Accuracy:
±
(2
%
of reading
+
20 V)