During the Discharge time, the display will show:
If the capacitance of the DUT is very small it is possible that you will not see this
display due to the short amount of time required to discharge the DUT.
The voltage and time of discharging DUT will be displayed during the discharge period
but this information is only intended for reference and cannot be reviewed after the
end of the discharge period. The process signal of the remote I/O remains active
during the discharge period and the high voltage lightning bolt indicator on the front
panel of the instrument is also active during the discharge period.
As a safety precaution, if the RESET button is pressed during the discharge period,
the instrument will log the RESET but will not discontinue the discharge function.
Pressing the RESET during the discharge will only prevent a connected memory from
executing after the discharge time.
The capacitance of the DUT should never exceed the maximum specification listed in
this manual.
DO NOT TOUCH THE DEVICE UNDER TEST, HIGH VOLTAGE
TEST LEAD, ALLIGATOR CLIP, OR CLIP INSULATOR
DURING THE DISCHARGE CYCLE.
5.3.2. Connected Memory indicator
The connected memory indicator is an underscore symbol located directly next to the
Memory number and only will be displayed when the connect function has been
turned ON in the Test parameters.
5.3.3. Manual Voltage Adjustment
When HYPOTMAX is performing a test, the
∧
and
∨
keys may be used to adjust the
output voltage.