62
COMMAND
NAME
DESCRIPTION
*PSC {1|0}
Power-on Status Clear Command 1 = Power-on clear enable
registers
0 = Power-on load previous enable
registers
*PSC?
Power-on Status Clear Query
*ESR?
Standard Event Status Register
Query
0 - 255
*ESE <
value
> Standard Event Status Enable
Command
value
= 0 - 255
*ESE?
Standard Event Status Enable
Query
0 - 255
*STB?
Read Status Byte Query
Read Status Byte
*SRE
<
value
>
Service Request Enable Command
value
= 0 - 255
*SRE?
Service Request Enable Query
0 - 255
*IDN?
Read the instrument identification string. Company =Associated Research Inc.
*RST
Reset the instrument to original power on configuration. Does not clear Enable
register for Standard Summary Status or Standard Event Registers. Does not clear
the output queue. Does not clear the power-on-status-clear flag.
*TST?
Performs a self-test of the instrument data memory. Returns 0 if it is successful or 1 if
the test fails.
*CLS
Clears the Status Byte summary register and event registers. Does not clear the
Enable registers.
*OPC
Sets the operation complete bit (bit 0) in the Standard Event register after a command
is completed successfully.
*OPC?
Returns an ASCII "1" after the command is executed.
*WAI
After the command is executed, it prevents the instrument from executing any further
query or commands until the no-operation-pending flag is TRUE.