72
Use the Numeric keys to enter the standard value of current, unit mA, and then press
the ENTER key to store the standard value of AC 20.00mA range for calibration and
the display will show:
AC Hipot, 3.500mA current range
Press the
∨
key, the program will advance to the AC 3.500mA range calibration of the
Hipot test. The display will show:
Connect a resistor about 100K
Ω
/40W in series with a AC standard Ammeter which
can measure up to 3.500mA to the output leads. The Ammeter should be connected
to the return lead, then press the TEST button, the program will automatically
generate an output on the output connectors about 300V/3.00mA and the display will
show:
Use the Numeric keys to enter the standard value of current, unit mA, and then press
the ENTER key to store the standard value of AC 3.500mA range of the hipot test and
the display will show: