43
Pressing the
∧
key will increase the output voltage, pressing the
∨
key will decrease
the output voltage. Manual voltage adjustment temporarily overrides the voltage
setting and only remains in effect until the test is terminated by Pass, Fail, or Abort.
When the Lock parameter is selected ON in the system menu, manual voltage
adjustment is disabled.
5.4. Reviewing Test Results
The method for reviewing test results is to press the ENTER key at the end of the test
sequence. Results can be reviewed at any time before the next test is executed by
pressing the ENTER key and then scrolling with the up and down arrow keys or
repeatedly pressing the ENTER key. The results of the last test in the process will be
followed by the first test when scrolling through the results. All buffers are cleared at
the start of the next test cycle.
Pressing the EXIT or RESET key will return you to the Settings screen.
5.5. Test Messages
If the test in process is Aborted with the RESET button or remote control, the display
will show:
If the test in process is Aborted with the RESET button or remote control before the
meter readings are taken, the display will show:
At the beginning of AC Withstand Voltage test when the voltage begins to ramp but
before the meter readings are taken, the display will show: