____________________________________________________________________________
71
Use the Numeric keys to enter the standard value of voltage, unit V, and then press
the ENTER key to store the standard value of AC or DC Hipot voltage for calibration
and display will show:
9.3.2. Calibration of 7705 AC Hipot Current
AC Hipot, 20mA current range
Press the
∨
key, the program will advance to the AC 20.00mA range calibration of the
Hipot test. The display will show:
Connect a resistor about 100K
Ω
/40W in series with a AC standard Ammeter which
can measure up to 20.00mA to the output leads. The Ammeter should be connected
to the return lead, then press the TEST button, the program will automatically
generate an output of about 2000V/20mA and the display will show: