70
7710,7720 1M
Ω
, 0.25-watt resistor, 300 volt
9.2. Calibration Initialization
Press and hold the calibration key on the rear panel with a pen, pencil, or small
screwdriver while powering ON the HypotMAX. The HypotMAX enters calibration
mode after the power on sequence is complete.
The program will automatically enter to the calibration mode and the display screen
will show:
Use the
∨
(Forward) and
∧
(Backward) keys to move through the calibration menu.
The following instructions follow the sequence scrolling forward through the menu.
9.3. Calibration Procedure
9.3.1. Calibration of Hipot Voltage
The first screen to appear in the Calibration mode is the Voltage calibration display.
The display will show:
Connect an AC standard voltage meter for 7705 and 7715 or a DC standard voltage
meter for 7710 and 7720 which can measure up to 10KV to the output connectors
and then press the TEST button, the program will automatically generate an output of
about 10KV and the display will show: