4. PROGRAMMING INSTRUCTIONS
4.1. Power Up
Turn on the POWER switch located on the lower left hand side of the front panel.
Upon powering the instrument up, a POWER ON SELF TEST will be automatically
performed. This test will check for the condition of the instrument’
s critical
components. In addition the display will show the following message, with the actual
model number and software version number:
The program will then recall the Memory which was last executed.
4.2. System Parameter Settings
Press the SETUP key to progress through the menu of System Parameters.
Successive key presses will advance the menu forward. The sequential forward
menu items are: PLC Remote, Fail Stop, Contrast, Alarm, Smart GFI, GPIB address,
BUS Remote Command Set, and Memory Lock. The system settings are global and
are not specific to any memory location.
4.2.1. PLC Remote
Press the SETUP key to advance to the PLC Remote parameter. The display will
show:
Use the ENTER key to select the mode PLC Remote Control.
Press the SETUP key to advance to the address setting if the GPIB interface card is
installed on this instrument. If this instrument does not have the GPIB interface card
(the program will advance to the contrast setting), or press the EXIT key to exit from
the PLC Remote selection to the operation mode. The instrument will store the
selection of PLC Remote automatically.