____________________________________________________________________________
57
SAO
Set the offset for the Continuity test. The cables and any test fixture should be
connected before executing the command. This command will perform an actual test
and all safety precautions should be observed when using this command.
SACG
Set the Charge-LO parameter for the DCW or IR test. The cables and any test fixture
should be connected before executing the command. The test parameters that are
set for the step will be used when performing the auto setting. This command will
perform an actual test and all safety precautions should be observed when using this
command.
7.4.3. File Editing Commands
The following commands are used to create or modify Test Setup Files.
COMMAND
DESCRIPTION
VALUE
FL <
memory
number
>
File Load
memory number
= 1-50
ADD
<
test
,
p1,p2,p3… >
Add all parameters
of a test
SF {1|0}
Fail Stop
1=On, 0=Off
FL <memory number>
Load a file by memory number from non-volatile memory into random access memory
RAM.
ADD <test,p1,p2,p3… >
This command edits all parameters in a memory location. Parameters will be edited
at the memory location that has been selected.
The parameter <
test>
indicates the test type. The test type values ACW or DCW
must be used. The parameters <
p1,p2>
etc. indicate the individual settings for each
test. All parameters must be included with the command and should appear in the
same order that is shown in the table below. Also, like the individual parameter
editing commands, the unit should not be included with the value. Only the numeric
value should be included in the command string. The list of parameters can also be
found in the default parameters section of the manual, or refer to the Test Parameter
Editing commands section for the proper values.
The parameter values should use complete text and not use the coded values that are
associated with the individual parameter setting commands such as "ON" and "OFF”.
The LS? companion command will also list all parameters in complete text in the order
as they appear in the following table, preceded by the memory number.