____________________________________________________________________________
19
Arc Detection
Range:
1 – 9
Failure Detector
Audible and Visual
Voltage Display
Range:
0.00 – 20.00 kV Full Scale
Resolution:
10 volt
Accuracy:
±
(2
%
of reading
+
20 V)
Current Display
Auto Range
Range 1:
0.0
µ
A - 350.0
µ
A
Resolution:
0.1
µ
A
Range 2:
300
µ
A - 5000
µ
A
Resolution:
1
µ
A
Accuracy:
±
(2
%
of reading
+
3 counts)
DC Output Ripple
≤
5
%
Ripple RMS at 20 kV DC @ 4999 uA, Resistive
Load
Output Regulation
±
(1
%
of setting
+
10 volts) from no load to full load
Maximum Capacitive Load 1.5 uF < 2 kV
0.28 uF < 4 kV
0.18 uF < 6 kV
0.15 uF < 10 kV
0.12 uF < 14 kV
0.10 uF < 16 kV
0.08 uF < 20 kV
Discharge time
No load
≤
500 ms. Capacitive loading will increase the
discharge time. The time will vary due to the load
condition and voltage selection.
Dwell Timer
Range:
0.0, 0.4 – 999.9 seconds
0.0 - 999.9 minutes (0 = Constant)
Resolution:
0.1 second or minute increments
Accuracy:
±
(0.1% + 1 count)
Ramp-UP
Range:
0.4 – 999.9 sec
Ramp-DOWN
Range:
0.0, 1.0 – 999.9 sec (0=OFF)
Resolution:
0.1 sec increments
Accuracy:
±
(0.1% + 1 count)
Ground Fault
GFI Trip
Current:
1 mA max
Interrupt
HV Shut Down
response:
< 1 ms (Actual Output Voltage discharge
time is not included and vary due to load
condition)
Содержание HypotMAX 7705