2.5.2 Valo leD curing light (if equipped)
cAUTiOn
– Refer to complete Manufacturer’s instructions for warnings
and proper use. The following is quick operation steps only.
To use, select the desired mode by pressing the lower Time/Mode Change button for 2 seconds and then releasing. The
Mode/Status Light will indicate the mode by color as follows:
Green – Standard Power
Orange – High Power
Flashing Orange – Xtra Power
See Manufacturer’s instructions for power selection.
Press the upper Power button to activate the light for curing. Press the power button again if it is desired to turn off the light
prior to it reaching its time cycle.
To change desired time intervals, press the Time/Mode button quickly to change intervals of time between 5, 10, 15 and 20
seconds.
2.6 Oral evacuation System
How it Works
To use an evacuation handpiece, remove the valve from the holder and insert the desired tip into the valve. The saliva ejector
valve accommodates standard disposable saliva ejector tips and the HVE valve accommodates standard disposable 7/8” tips.
To ensure maximum suction flow, periodically check and replace the solids trap screen located on the front of the system.
Remove the cover to the solids trap, lift out the old screen, insert the new screen and replace the cover to the solids trap.
2.7 Self-contained Vacuum Pump
The vacuum pump is activated when the Suction/Vacuum On/Off switch, located on the rear of the unit, is in the ON
position, and a suction handpiece is removed from the holder.
To use, open the lever on the evacuation handpiece. To maximize vacuum strength of the evacuation handpiece, close the
lever valve of the other evacuation handpiece whenever it is not in use.
NOTE: The strength of the vacuum pull to the oral evacuation system is regulated by the vacuum relief valve. The relief
valve on the system is pre-set to approximately 12” Hg, which is ideal for most general dental as well as surgical
procedures.
2.7.1 Sediment Separator and Holding Tank
cAUTiOn:
The self-contained suction system includes a suction canister for holding liquid contents obtained from
evacuation by the suction handpieces. The contents must be emptied daily. The suction canister has an overflow
switch to help prevent accidental overfilling of the canister. However it is important to empty the liquid before it
reaches capacity to prevent potential liquid being pulled into the vacuum pump. Follow the instructions under maintenance for
purging and cleaning the suction system.
The bottom of the suction canister has a sediment bowl which is designed to provide an internal separator. The system uses the
settling method for heavy particles including fine particles of dental filling amalgam which may be captured in the sediment
bowl. The sediment bowl should be checked every three months. See Maintenance instructions for checking and replacing the
Sediment Separator bowl.
Valo LED Curing Light